    高低溫試驗箱工作室的參數不僅關系到室內氣流的流速,同時也關系到可容納的樣品體積。對高低溫試驗箱工作室的要求有:

    (1)高(gao)(gao)(gao)低(di)(di)溫(wen)試(shi)(shi)(shi)驗(yan)(yan)(yan)箱(xiang)相對樣(yang)(yang)品應(ying)有足夠(gou)的(de)容(rong)量(liang),以保證每個受(shou)試(shi)(shi)(shi)的(de)樣(yang)(yang)品處于(yu)相同的(de)環境條件(jian)(jian)下,并保證試(shi)(shi)(shi)驗(yan)(yan)(yan)時(shi)樣(yang)(yang)品不(bu)干擾試(shi)(shi)(shi)驗(yan)(yan)(yan)條件(jian)(jian)。為使(shi)高(gao)(gao)(gao)低(di)(di)溫(wen)試(shi)(shi)(shi)驗(yan)(yan)(yan)箱(xiang)室中的(de)試(shi)(shi)(shi)驗(yan)(yan)(yan)條件(jian)(jian)能模擬實際(ji)使(shi)用條件(jian)(jian),高(gao)(gao)(gao)低(di)(di)溫(wen)試(shi)(shi)(shi)驗(yan)(yan)(yan)箱(xiang)室的(de)體積(ji)(ji)與試(shi)(shi)(shi)驗(yan)(yan)(yan)樣(yang)(yang)品體積(ji)(ji)之比應(ying)不(bu)小于(yu)5:1。對工作時(shi)發(fa)熱量(liang)很大(da)的(de)產品,如電機和變壓器等,高(gao)(gao)(gao)低(di)(di)溫(wen)試(shi)(shi)(shi)驗(yan)(yan)(yan)箱(xiang)對試(shi)(shi)(shi)驗(yan)(yan)(yan)樣(yang)(yang)品的(de)體積(ji)(ji)比應(ying)不(bu)小于(yu)10:1。

高低溫試驗箱工作室、樣品測試及安裝要求

    (2)高低溫試驗箱室應能保證試驗期間的環境參數精度,并定期或在每一試驗前后進行鑒定。為了滿足試驗要求,用于控制箱溫的溫度傳感器應盡可能位于室的,并加以遮蓋或保護,以防止輻射熱造成誤差。
    (3)除考慮室壁輻射熱和漏入的熱量是試驗控制條件外,通常應使高低溫試驗箱室的熱源或冷源的輻射熱不直接落到樣品上,以排除測試樣品時的干擾因素。
    當對同一樣品依次進行兩種或兩種以上試驗時,試驗順序的編排對試驗后果有很大的影響。通常遵循的原則是:
    (1)按產品在運輸、使用中所遇到的主要環境影響的順序進行環境模擬試驗。
    (2)如果希望從一個樣品上獲取盡可能多的環境適應性信息,則應把對影響較輕的試驗,如溫度試驗、低氣壓試驗等排在前面。
    (3)如果要求盡快暴露產品缺陷,應使前一試驗對樣品的影響在后一試驗中得到積累和加強,可排為高溫試驗、振動試驗、沖擊試驗或跌落試驗、濕熱試驗、低溫試驗的順序。
    (4)從經濟角度考慮,試驗周期長和花費大的試驗項目應排在后。
    試驗樣品的安裝位置直接關系到熱輻射對工作室內溫度場分布的影響,安裝位置通常有以下原則:
    (1)試品應盡可能地靠近工作室的中央,使試品的任何部分(包括安裝架)和室壁之間有盡可能大的空間,小距離應大于等于10cm,以利于氣流的流動和溫度的均勻。
    (2)對于形狀不規則的樣品,其擺放應按有利于氣流流通的原則。并且盡量使氣流的流通截面平穩,以不造成氣流組織流動的紊亂。
    (3)對于散熱型的樣品,應將散熱較大的面對向氣流流動的地方,以使散熱面和氣流進行充分熱交換。防止樣品的散熱不均影響工作室內溫度的均勻。